modernmicroscopy : columns : "how to" tutorial series


How to Obtain the Best Resolution With a Modest Light Microscope
by  Walter Franchini

4/10/20121 | 2 | 3 | NEXT>> 

One of the most important tests to verify the potential of a light microscope is to use a diatom slide of Amphipleura pellucida (Kützing) mounted in a high refractive index medium, such as Zrax (n=1.70+) (Figure 1). This type of diatom lives in freshwater lakes and rivers. From the Victorian era to the present day, separation of the lines (striae) and dots (punctae) of this diatom has always been a challenge for the microscopist. The striae of this diatom have a separation of 250 nm (40 to 42 lines in 10 µm; 50 to 52 dots in 10 µm) which is near the limit of optical resolution of a modest light microscope.

 

click image to enlarge (188K)

FIGURE 1

 

To achieve the separation of the striae and dots you must have accurate control over every single element of the microscope, including, first of all, the objective and condenser of the microscope. You can use a high power objective such as the 63X or 100X with a numerical aperture of at least 1.25, and possibly of the fluorite or apochromatic type with high correction for chromatic and spherical aberration (Figure 2).

 

click image to enlarge (194K)

FIGURE 2

 

The substage condenser must have the characteristics of N.A. ≥ 1.25. A very good condenser of this type is the old Zeiss or LOMO aplanatic condenser with numerical aperture of 1.4 whose aperture diaphragm is on rack and pinion, and rotatable for obtaining oblique illumination from any azimuth (Figure 3).

 

click image to enlarge (126K)

FIGURE 3

 


4/10/20121 | 2 | 3 | NEXT>>