| |
modernmicroscopy
:
articles
Contaminant Identification in Pharmaceutical Products
| by |
Gretchen L. Shearer, Ph.D., McCrone Associates, Westmont, IL |
Examples of
Contamination
Fibers
Fibers are a commonly
encountered contaminant. Fibers are usually identified using polarized
light microscopy alone, but FTIR is sometimes used to confirm the chemical
identification. Sometimes, SEM is used to examine the fiber morphology.
Most often, paper, cotton, and colorless polyester fibers are observed
in pharmaceuticals. These fibers are related to cleanroom wipes and garments.
Sometimes, a variety of fibers is found within the same sample, and this
is an indication of a more serious contamination issue. Figure
1 shows some typical fibers recovered from a pharmaceutical product.
|
click image to enlarge (228K)
FIGURE
1
Examples of paper, cotton and
polyester fibers recovered
from a pharmaceutical product.
|
Glass
Glass is another contaminant
that is often observed during quality control screening. Glass particles can
be generated by fracture of the vial neck or opening, from external sources
such as other vials, glassware and lighting, and from delamination of the inner
vial surface. Because of their density, glass particles sink rapidly to the
bottom of the vial when the liquid is agitated. Glass delamination flakes are
extremely thin and may be missed during visual examination. If a larger number
are present, a “twinkling” effect in the solution is observed. The “twinkling”
is best observed using a fiber optic light source. Delamination flakes do not
sink to the bottom of the vial. Glass delamination usually occurs when a
highly acidic, highly basic or sodium chloride solution is stored in the wrong
type of container.
|
click image to enlarge (164K)
FIGURE
2
Glass particles recovered from
a
pharmaceutical product, as they
appear with oblique illumination.
|
Glass particles have
distinctive morphology and may be identified by microscopical analysis.
Figure 2 shows several glass particles on a polycarbonate
membrane filter. The glass particles are shiny, refractive and have a
characteristic conchoidal fracture. When examined using polarized light,
glass is isotropic. Refractive index measurement and elemental analysis
can be used to determine the type of glass and to compare it to suspect
sources. The elements lithium and boron are present in some types of
glass, and these light elements can not be detected using EDS detectors;
if detection of these elements is needed, the particles can be analyzed
using secondary ion mass spectrometry (SIMS).
| click image to enlarge (224K) 
FIGURE
3
Glass delamination flakes
recovered from a pharmaceutical
product, as they appear on a
polycarbonate filter membrane.
Note the small holes or pits on the
flakes. Image was taken using
episcopic illumination.
|
click image to enlarge (301K)
FIGURE
4
Glass delamination on inside of
vial wall.
This image shows the delamination at the
bottom of the vial. Image was taken using
episcopic illumination.
|
click image to enlarge (290K)
FIGURE
5
Glass delamination on inside of vial
wall.
The fill line of the vial is evident. Image was
taken using episcopic illumination. |
Glass delamination
flakes as they appear on the filter membrane are shown in Figure
3. Using correct illumination is critical. Due to their extreme
thinness, the flakes would not be observed on a membrane filter using
oblique or side light; they are only seen using episcopic illumination
or top light. If glass delamination flakes are detected, the interior
of the sample container should also be examined microscopically. The
beginnings of delamination are observed as small pitted areas resembling
circles and doughnuts on the interior glass surface. A vial interior exhibiting
a severe case of delamination is presented in Figure 4.
The heaviest delamination tends to occur at the bottom sides of the vial,
and lessens further up the vial sides. In Figure 5,
the fill line of the vial is evident.
Silicone
Silicone is used
in many products and it is frequently found in pharmaceutical products.
When it is used as a lubricant for rubber stoppers and plungers, it
very easily “sloughs off” and ends up in the product. Silicone can
interact with protein based drugs or active ingredients producing particles
or residue. Thermal degradation of the silicone can occur if the product
is autoclaved. Silicone oil can be observed in liquid products as oil
droplets. Sometimes it occurs as fine droplets that give the solution
a hazy appearance. After filtration of the product, silicone oil appears
as “cleared” areas on the filter that can be observed using transmitted
light. Figure 6 shows the cleared zones on a polycarbonate
filter. Degraded silicone, or silicone that has interacted with protein
or active ingredient, forms a semi-solid particulate residue that appears
as stringy “fiber-like” particles in solution. Stringy silicone particulate,
as it appears on the membrane, is shown in Figure 7.
It is essential to use the correct type of illumination (epi-illumination
in this case), as this type of residue often cannot be seen with transmitted
or oblique illumination.
click image to enlarge (271K)
FIGURE
6
Cleared areas on a polycarbonate
filter
membrane caused by silicone. Image was
taken using transmitted light. |
click image to enlarge (348K)
FIGURE
7
Stringy, degraded silicone particulate
residue as it appears on a polycarbonate
filter membrane. The image was taken using
episcopic illumination. |
Suspect silicone
residues are prepared for analysis using FTIR to confirm the identification.
The silicone oil can be recovered from the filters with oily or cleared
zones by extracting a small portion of the filter with nonane. This
procedure is performed onto a polished salt plate for FTIR analysis.
Stringy residues can be scraped off the filter using a fine tungsten
needle and prepared for FTIR. Small amounts of silicone residue can
be prepared for analysis using special replication techniques. Infrared
spectra of silicone oil and stringy silicone particulate from pharmaceutical
products are compared in Figure 8, and a typical spectrum
of silicone/protein residue is presented in Figure 9.
click image to enlarge (98K)
FIGURE
8
FTIR spectrum of (top) silicone
oil
extracted from a polycarbonate filter,
(middle) stringy degraded silicone
particulate residue, and (bottom) a silicone
reference. |
click image to enlarge (63K)
FIGURE
9
FTIR spectrum of silicone/protein
residue. |
Vials
Particles and residues
can occur as a result of processing and sometimes are observed on the
vials prior to filling. Figure 10 shows a residue
from an unfilled vial. FTIR analysis indicated that the material was
similar to the processing detergent (Figure 11).
click image to enlarge (49K)
FIGURE
10
Light brown residue in an unfilled
vial. |
click image to enlarge (78K)
FIGURE
11
FTIR spectrum of the light brown
residue
from an unfilled vial (Figure 10) compared to
a spectrum of the processing detergent.
|
Occasionally, a
vial will be submitted that was thought to contain a dark particle.
Optical examination indicates that the defect or particle is, in fact,
embedded in the glass or is contained entirely within the wall of the
vial. Figure 12 shows black material that is embedded
in the vial glass. The material was identified as hematite (iron oxide
or rust) using Raman microscopy (Figure 13). The presence
of iron and oxygen can be confirmed using EDS. Raman has the benefit
of giving compound information that complements the elemental data.
click image to enlarge (61K)
FIGURE
12
Black material embedded in glass
vial
wall. |
click image to enlarge (78K)
FIGURE
13
Raman spectrum of the black material
embedded in glass vial wall (Figure 12)
compared to a Raman spectrum of hematite
(iron oxide).
|
|