| |
modernmicroscopy
:
articles
Evaluation of a Prototype BF-DF-Oblique-Circular Oblique Lighting (BF-DF-Obl-COL) Condenser
| by |
Ted Clarke, Scientific Photographer and Instrument Maker |
Imaging the Diatom Test Plate with the 4X Objective
Figure 8 shows the condenser in the lowered position
with the top lens removed for use with the 4X Zeiss objective. (I now
use a Zeiss 4X 160 mm tube length objective instead of the LOMO 4X objective
because the Zeiss lens is corrected for use with a compensating Zeiss
Kpl eyepiece found best for use with the higher power LOMO objectives.)
Figure 9 shows the diatoms imaged in brightfield with the 4X objective
with the field diaphragm adjusted so its image falls just within the 18
mm diameter intermediate image field size of the stop in the 10X Zeiss
Kpl high eyepoint eyepiece. Figure 10 shows the same field imaged in darkfield
after selecting and centering the stop in the wheel and opening the aperture
diaphragm in the condenser. This low magnification is important for surveying
the field before switching to a higher power objective.
click image to enlarge (191K)
Figure
8 |
click image to enlarge (114K)
Figure
9 |
click image to enlarge (135K)
Figure
10 |
Imaging Pleurosigma angulatum with Darkfield Illumination
The diatom test slide has the diatom Pleurosigma angulatum
which serves as a very good resolution test target for the 40X 0.65 NA
objective because the stria spacing of about 0.52 micrometers matches
the theoretical resolution of a 0.65 NA objective, when used with a matching
illumination NA of just under 0.65. In his response to a letter by Robert
B. McLaughlin,4 Dr. Walter McCrone republished a very high
resolution optical photomicrograph of Pleurosigma angulatum taken
in the early 20th Century by Spitta. This image is shown in
Figure 11. It was evidently taken with darkfield illumination, probably
with blue light and perhaps with even shorter wavelength ultraviolet. The
definition of “just resolved” means that the periodicity of the structure
will be detectable but this fine structure will not be faithfully resolved.
Figure 12 shows Pleurosigma angulatum recorded with the LOMO 40X
objective and darkfield illumination from the prototype condenser. The
condenser height had to be raised from the brightfield setting in order
for the high NA rays to illuminate the specimen for darkfield. This same
problem exists with the Olympus Abbe condenser used with my darkfield
inserts and will be discussed in a subsequent paragraph. The CoolPix®
lens zoom control was set so the diagonals of the recorded field covered
the field seen with the 10X 18 mm FN eyepiece and the image was subsequently
cropped in Adobe PhotoShop®. The periodicity is recorded along
with lines that initially were suspicious of being alias lines from the
camera sensor resolution being close to the optical resolution. The CoolPix®
lens was zoomed to cover about half the field size of the eyepiece for
the cropped portion of the field shown in Figure 13. The lines are still
present and therefore not from digital camera aliasing because they are
also evident on a close examination through the eyepiece. Tony Havics
of pH2, LLC previously tested my modified Biolam. Tony found that darkfield,
with the stop at the fiber-optic light guide end with the aplanatic condenser
and the same 40X objective, was capable of resolving the first three sets
of lines of the HSE/NPL Test Slide, as shown in Figure 14. Resolving the
three sets of lines is a requirement for counting asbestos fibers using
phase contrast microscopy.
Figure
11 |
click image to enlarge (163K)
Figure
12 |
click image to enlarge (215K)
Figure
13 |
click image to enlarge (164K)
Figure
14 |
|